Reliability of a measurement method for the cross-sectional area of the longus colli using real-time ultrasound imaging
被引:4
|作者:
O'Riordan, Cliona
[1
]
Van De Ven, Pepijn
[2
]
Nelson, John
[2
]
McCreesh, Karen
[1
]
Clifford, Amanda
[1
]
机构:
[1] Univ Limerick, Dept Clin Therapies, Hlth Sci Bldg,2nd Floor, Limerick, Ireland
[2] Univ Limerick, Dept Sci & Engn, Limerick, Ireland
来源:
关键词:
Real-time;
ultrasound;
longus colli;
reliability;
D O I:
10.1177/1742271X16659099
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Objective Real-time ultrasound imaging is an established objective outcome measurement with proven reliability. However, it is still largely biased by user-ability. Published research in the area of real-time ultrasound imaging reliability in the cervical region and in particular the deep cervical flexors is quite sparse. The purpose of this investigation was to examine if a novice ultrasound user could agree favourably with an experienced ultrasound sonographer in measuring the cross-sectional area of the longus colli. Methods Ultrasound images were captured from 22 healthy subjects on two different occasions, one week apart, by a novice ultrasound user. They were acquired using a GE Healthcare LOGIQe ultrasound machine, at a depth of 3-4cm with the transducer frequency set to 8MHz, in line with previous research guidelines. Cross-sectional area was then measured on-screen by both a novice and experienced ultrasound user to determine inter-rater reliability. Intra-rater reliability was also analysed using cross-sectional area figures from days one and two. Results Intra-rater reliability for real-time ultrasound imaging for the cross-sectional area of the longus colli was excellent (intra-class correlation 0.90, 95% CI 0.82-0.95). Inter-rater reliability was moderate but in keeping with previous published research (intra-class correlation 0.61, 95% CI 0.37-0.77). Conclusion Difficulties in identifying the borders of the longus colli muscle due to its deep anatomical location and surrounding structures make it difficult for assessors to agree favourably on cross-sectional area measurements, leading to moderate levels of inter-rater reliability and poor agreement. Intra-rater reliability is excellent, and in this instance indicates that a novice user can be just as reliable as a more experienced ultrasound user.
引用
收藏
页码:154 / 162
页数:9
相关论文