PATHOLOGICAL FIBRIN FORMATION AND COLD-INDUCED CLOTTING OF MEMBRANE OXYGENATORS DURING CARDIOPULMONARY BYPASS

被引:32
作者
BLOMBACK, M
KRONLUND, P
ABERG, B
FATAH, K
HANSSON, LO
EGBERG, N
MOOR, E
CARLSSON, K
机构
[1] KAROLINSKA HOSP, DEPT THORAC SURG, S-10401 STOCKHOLM, SWEDEN
[2] KAROLINSKA HOSP, DEPT MED CARDIOL, S-10401 STOCKHOLM, SWEDEN
[3] ROYAL INST TECHNOL, S-10044 STOCKHOLM, SWEDEN
关键词
CARDIOPULMONARY BYPASS; COLD-INDUCED CLOTTING; OXYGENATORS; FIBRIN GEL NETWORK;
D O I
10.1016/S1053-0770(05)80053-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In 1,800 patients undergoing cardiac surgery over a P-year period, 11 incidents of abnormal inlet pressure elevations occurred before the membrane oxygenators. In 3 patients, the oxygenators had to be changed during cardiopulmonary bypass. This complication was found to be caused by fibrin formation possibly secondary to precipitaition of fibrinogen with other coagulation factors in the heat exchangers of the oxygenators during the cooling phase. Large amounts of fibrin were demonstrated in the heat exchanger of the oxygenators. After careful washing of the apparatus, plasmin was added and fibrin was detected by measuring D-dimer levels. In heat exchangers from uneventful operations, only trace amounts of fibrin were found. Because there were no cold agglutinins demonstrated in the patients before surgery, cryoprecipitation studies were performed soon after surgery. When the patients' plasma samples were studied at different temperatures, from 37 degrees C down to 3 degrees C, cryoprecipitates or a gel (in 1 patient only) were formed. This indicated that there might be something abnormal with regard to fibrinogen fibrin formation. The study patients were therefore investigated after the acute phase of the operation had ended for various coagulation factors, as well as for fibrin gel network characteristics. The results were compared with those of a control group (n = 10) with uneventful operations. There were no differences between the groups with regard to levels of coagulation factors VII and VIII and von Willebrand factor, although they were increased in both groups. The mean levels of coagulation inhibitors, antithrombin and Protein S, were slightly lower in the study patients. All of these patients had a highly pathologic, ie, tight fibrin gel network, except for the patient in whose sample a gel formed, despite being treated with aspirin or oral anticoagulants. The network was also tighter in some of the controls (v middle-aged reference individuals), although it was significantly tighter in the patients. It is concluded that some individuals who have an increased tendency to form tighter fibrin gel networks might be at increased risk for a severe complication during cardiac surgery performed under hypothermia. Copyright (C) 1995 by W.B. Saunders Company
引用
收藏
页码:34 / 43
页数:10
相关论文
共 40 条
[1]   ANTITHROMBIN (HEPARIN COFACTOR) ASSAY WITH NEW CHROMOGENIC SUBSTRATES (S-2238 AND CHROMOZYM-TH) [J].
ABILDGAARD, U ;
LIE, M ;
ODEGARD, OR .
THROMBOSIS RESEARCH, 1977, 11 (04) :549-553
[2]  
BJORNSSON TD, 1989, J PHARMACOL EXP THER, V250, P154
[3]   NATIVE FIBRIN GEL NETWORKS OBSERVED BY 3D MICROSCOPY, PERMEATION AND TURBIDITY [J].
BLOMBACK, B ;
CARLSSON, K ;
HESSEL, B ;
LILJEBORG, A ;
PROCYK, R ;
ASLUND, N .
BIOCHIMICA ET BIOPHYSICA ACTA, 1989, 997 (1-2) :96-110
[4]  
BLOMBACK M, 1983, PEDIATRICS, V72, P416
[5]   THE EFFECTS OF PREOPERATIVE ASPIRIN THERAPY ON PLATELET-FUNCTION IN CARDIAC-SURGERY [J].
BOLDT, J ;
KNOTHE, C ;
ZICKMANN, B ;
HEROLD, C ;
DAPPER, F ;
HEMPELMANN, G .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (11) :598-602
[6]  
BULL BS, 1975, J THORAC CARDIOV SUR, V69, P674
[8]   LACK OF CORRELATION BETWEEN ACTIVATED CLOTTING TIME AND PLASMA HEPARIN DURING CARDIOPULMONARY BYPASS [J].
CULLIFORD, AT ;
GITEL, SN ;
STARR, N ;
THOMAS, ST ;
BAUMANN, FG ;
WESSLER, S ;
SPENCER, FC .
ANNALS OF SURGERY, 1981, 193 (01) :105-111
[9]  
DECLERCK PJ, 1988, BLOOD, V71, P220
[10]  
Dellora Carlo, 1992, Perfusion, V7, P195, DOI 10.1177/026765919200700303