A study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy

被引:31
作者
Rajasekaran, Dhanabalan [1 ]
Jeevanandam, Prakash [1 ]
Sukumar, Prabakar [1 ]
Ranganathan, Arulpandiyan [1 ]
Johnjothi, Samdevakumar [1 ]
Nagarajan, Vivekanandan [1 ]
机构
[1] Canc Inst WIA, Dept Med Phys, Madras, Tamil Nadu, India
关键词
Octavius 4D system; VMAT QA; 3D dosimetry; Dose verification; Plan complexity;
D O I
10.1016/j.rpor.2014.08.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: To evaluate the new Octavius 4D system for patient specific quality assurance and to study the correlation between plan complexity and gamma index analysis in patient specific quality assurance of VMAT using the Octavius 4D system. Background: McNiven (2010) proposed a study to evaluate the utility of a complexity metric, the Modulation Complexity Score, to evaluate the relationship of the metric with deliver-ability in IMRT. Materials and methods: Evaluation of the Octavius 4D system was carried out by gamma evaluation of user defined MLC created patterns and AAPM TG 119 benchmark plans. The relationship between plan complexity expressed as Modulation Complexity Score (MCS) and the gamma index analysis was established by a planar and volumetric gamma analysis of 106 clinically approved VMAT patient plans of different sites. Results: Average volumetric 3D global gamma evaluation (3 mm/3%) results for the evaluation plans was 97.41% for 6 MV X-rays and 98.30% for 15 MV X-rays. Average MCS values for the head and neck, pelvic and thoracic plans were 0.2224, 0.3615 and 0.1874. Average volumetric 3D global gamma analysis (3 mm/3%) results for the head and neck, pelvic and thoracic VMAT plans were 95.45%, 97.51% and 96.98%, respectively. Out of 90 correlation analyses between the MCS and gamma passing rate, only 3 had the r value greater than 0.5. Conclusions: The Octavius 4D system is a suitable device for patient specific pretreatment QA. Global and local gamma analysis results showed a weak correlation with the MCS. (C) 2014 Greater Poland Cancer Centre. Published by Elsevier Urban & Partner Sp. z o.o. All rights reserved.
引用
收藏
页码:57 / 65
页数:9
相关论文
共 15 条
[1]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[2]   Intensity-modulated radiotherapy: Current status and issues of interest [J].
Boyer, AL ;
Butler, EB ;
DiPetrillo, TA ;
Engler, MJ ;
Fraass, B ;
Grant, W ;
Ling, CC ;
Low, DA ;
Mackie, TR ;
Mohan, R ;
Purdy, JA ;
Roach, M ;
Rosenman, JG ;
Verhey, LJ ;
Wong, JW ;
Cumberlin, RL ;
Stone, H ;
Palta, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :880-914
[3]   Quantification of beam complexity in intensity-modulated radiation therapy treatment plans [J].
Du, Weiliang ;
Cho, Sang Hyun ;
Zhang, Xiaodong ;
Hoffman, Karen E. ;
Kudchadker, Rajat J. .
MEDICAL PHYSICS, 2014, 41 (02)
[4]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[5]   Evaluation of a new VMAT QA device, or the "X" and "O" array geometries [J].
Feygelman, Vladimir ;
Zhang, Geoffrey ;
Stevens, Craig ;
Nelms, Benjamin E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :146-168
[6]   Implementing IMRT in clinical practice: A joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in Medicine [J].
Galvin, JM ;
Ezzell, G ;
Eisbrauch, A ;
Yu, C ;
Butler, B ;
Xiao, Y ;
Rosen, I ;
Rosenman, J ;
Sharpe, M ;
Xing, L ;
Xia, P ;
Lomax, T ;
Low, DA ;
Palta, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (05) :1616-1634
[7]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661
[8]   The importance of accurate treatment planning, delivery, and dose verification [J].
Malicki, Julian .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2012, 17 (02) :63-65
[9]   Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy [J].
Masi, Laura ;
Doro, Raffaela ;
Favuzza, Virginia ;
Cipressi, Samantha ;
Livi, Lorenzo .
MEDICAL PHYSICS, 2013, 40 (07)
[10]   Octavius 4D characterization for flattened and flattening filter free rotational deliveries [J].
McGarry, Conor K. ;
O'Connell, Barry F. ;
Grattan, Mark W. D. ;
Agnew, Christina E. ;
Irvine, Denise M. ;
Hounsell, Alan R. .
MEDICAL PHYSICS, 2013, 40 (09)