INFLATION STATIC PRESSURE-VOLUME CURVES OF THE TOTAL RESPIRATORY SYSTEM DETERMINED WITHOUT ANY INSTRUMENTATION OTHER THAN THE MECHANICAL VENTILATOR

被引:25
作者
FERNANDEZ, R
BLANCH, L
ARTIGAS, A
机构
[1] Servicio de Medicina Intensiva, Hospital de Sabadell, Sabadell, E-08208, Parc Tauli s/n
关键词
PULMONARY FUNCTION TEST; PRESSURE-VOLUME CURVES; MECHANICAL VENTILATION; ACUTE RESPIRATORY FAILURE;
D O I
10.1007/BF01709275
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To assess the accuracy of static pressure-volume (PV) curves of the total respiratory system performed with the data directly obtained from the Servo 900C vs. the data obtained from an external calibrated device. Design: Performance of the PV curve by the method of Levy with simultaneous recording of data obtained from both systems. Setting: The general ICU of Hospital de Sabadell. Patients: Ten sedated and paralyzed patients ventilated in the control mode for acute respiratory failure were evaluated. Interventions and measurements: Inflation static PV curves were performed by the method of Levy. We simultaneously measured airway pressure and volume by means of calibrated pressure tranducer and pneumotachograph and by the internal devices built in the ventilator Siemens Servo 900C. Statistics were concordance analysis between the two methods and covariance analysis between derived curves. Results: concordance analysis between both methods showed a 95% confidence interval of (+4%, -5%) in volume and (+2.2 cmH2O, -1.7 cmH2O) in pressure. Derived PV curves analyzed by MANOVA showed no significant differences whichever the method used within subject (p = 0.579). Conclusion: Inflation static pressure volume curves of the total respiratory system can be accurately performed with the data directly obtained from the Servo Ventilator 900C without the need of any other external device.
引用
收藏
页码:33 / 38
页数:6
相关论文
共 20 条
[1]  
BENITO S, 1985, INTENS CARE MED, V11, P76
[2]   PULMONARY PRESSURE-VOLUME RELATIONSHIP IN ACUTE RESPIRATORY-DISTRESS SYNDROME IN ADULTS - ROLE OF POSITIVE END EXPIRATORY PRESSURE [J].
BENITO, S ;
LEMAIRE, F .
JOURNAL OF CRITICAL CARE, 1990, 5 (01) :27-34
[3]  
BENITO S, 1991, UPDATE INTENSIVE CAR, V13, P86
[4]   EFFECT OF PEEP ON THE ARTERIAL MINUS END-TIDAL CARBON-DIOXIDE GRADIENT [J].
BLANCH, L ;
FERNANDEZ, R ;
BENITO, S ;
MANCEBO, J ;
NET, A .
CHEST, 1987, 92 (03) :451-454
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   COMPLIANCE AND DYNAMIC CHARACTERISTICS CURVES IN ACUTE RESPIRATORY-FAILURE [J].
BONE, RC .
CRITICAL CARE MEDICINE, 1976, 4 (04) :173-179
[7]  
CAPPS J S, 1987, Respiratory Care, V32, P558
[8]   CAUSES OF ERROR OF RESPIRATORY PRESSURE-VOLUME CURVES IN PARALYZED SUBJECTS [J].
DALLAVASANTUCCI, J ;
ARMAGANIDIS, A ;
BRUNET, F ;
DHAINAUT, JF ;
CHELUCCI, GL ;
MONSALLIER, JF ;
LOCKHART, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (01) :42-49
[9]   MECHANICAL EFFECTS OF PEEP IN PATIENTS WITH ADULT RESPIRATORY-DISTRESS SYNDROME [J].
DALLAVASANTUCCI, J ;
ARMAGANIDIS, A ;
BRUNET, F ;
DHAINAUT, JF ;
NOUIRA, S ;
MORISSEAU, D ;
LOCKHART, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1990, 68 (03) :843-848
[10]   PULMONARY AND CHEST-WALL MECHANICS IN ANESTHETIZED PARALYZED HUMANS [J].
DANGELO, E ;
ROBATTO, FM ;
CALDERINI, E ;
TAVOLA, M ;
BONO, D ;
TORRI, G ;
MILICEMILI, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1991, 70 (06) :2602-2610