LARGE-VOLUME HEMOCYTOMETER CHAMBER FOR ACCURATE COUNTING OF WHITE CELLS (WBCS) IN WBC-REDUCED PLATELETS - VALIDATION AND APPLICATION FOR QUALITY-CONTROL OF WBC-REDUCED PLATELETS PREPARED BY APHERESIS AND FILTRATION

被引:72
作者
LUTZ, P [1 ]
DZIK, WH [1 ]
机构
[1] HARVARD UNIV,SCH MED,DEPT MED,BOSTON,MA 02115
关键词
D O I
10.1046/j.1537-2995.1993.33593255602.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A detailed description is provided of a method using a large-volume (50 muL) hemocytometer to count very low numbers of white cells (WBCs) in platelet components. A method employing a Nageotte hemocytometer uses crystal violet stain and a standard microscope with a reading time of 5 to 10 minutes. .The method is validated by using serial dilutions of known concentrations of WBCs in platelets and by correlation with a flow cytometric technique. The interobserver coefficient of variation was 11.9 percent for WBC concentrations >2 per muL. Use of the method for the evaluation of 203 WBC-reduced platelet components prepared by apheresis or by filtration revealed that over 94 percent of components had WBC content <5 x 10(6) This method could easily be applied in the routine quality control of WBC-reduced platelet components and in clinical studies employing these components.
引用
收藏
页码:409 / 412
页数:4
相关论文
共 12 条
[1]   A FLOW CYTOMETRIC TECHNIQUE TO ACCURATELY MEASURE POST-FILTRATION WHITE BLOOD-CELL COUNTS [J].
BODENSTEINER, DC .
TRANSFUSION, 1989, 29 (07) :651-653
[2]  
BURGSTALER E A, 1991, Transfusion (Bethesda), V31, p21S
[3]   SAMPLING ERRORS AND THE PRECISION ASSOCIATED WITH COUNTING VERY LOW NUMBERS OF WHITE CELLS IN BLOOD COMPONENTS [J].
DUMONT, LJ .
TRANSFUSION, 1991, 31 (05) :428-432
[4]   FLOW-CYTOMETRIC METHOD FOR COUNTING VERY LOW NUMBERS OF LEUKOCYTES IN PLATELET PRODUCTS [J].
DZIK, WH ;
RAGOSTA, A ;
CUSACK, WF .
VOX SANGUINIS, 1990, 59 (03) :153-159
[5]   WHITE CELL-REDUCED BLOOD COMPONENTS - SHOULD WE GO WITH THE FLOW [J].
DZIK, WH .
TRANSFUSION, 1991, 31 (09) :789-791
[6]   WHITE CELL COUNTING IN RED-CELLS AND PLATELETS - HOW FEW CAN WE COUNT [J].
FRIEDMAN, LI ;
SADOFF, BJ ;
STROMBERG, RR .
TRANSFUSION, 1990, 30 (05) :387-389
[7]   ACCURATE QUANTITATION OF THE LOW NUMBER OF WHITE CELLS IN WHITE CELL-DEPLETED BLOOD COMPONENTS [J].
KAO, KJ ;
SCORNIK, JC .
TRANSFUSION, 1989, 29 (09) :774-777
[8]  
KJELDSBERG CR, 1986, BODY FLUIDS LABORATO, P154
[9]   VALIDATION OF A SIMPLE METHOD TO COUNT VERY LOW WHITE CELL CONCENTRATIONS IN FILTERED RED-CELLS OR PLATELETS [J].
MASSE, M ;
NAEGELEN, C ;
PELLEGRINI, N ;
SEGIER, JM ;
MARPAUX, N ;
BEAUJEAN, F .
TRANSFUSION, 1992, 32 (06) :565-571
[10]   A MULTICENTER STUDY ON THE EFFICIENCY OF WHITE CELL REDUCTION BY FILTRATION OF RED-CELLS [J].
MASSE, M ;
ANDREU, G ;
ANGUE, M ;
BABAULT, C ;
BEAUJEAN, F ;
BIDET, ML ;
BOUDART, D ;
CALOT, JP ;
COTTE, C ;
FOLLEA, G ;
GEROTA, J ;
HAU, F ;
HUREL, C ;
LEGRAND, E ;
MARCHESSEAU, B ;
NASR, O ;
ROBERT, F ;
ROYER, D ;
SCHOONEMAN, F ;
TARDIVEL, R ;
VIDAL, M .
TRANSFUSION, 1991, 31 (09) :792-797