UTILITY OF AN END-TIDAL CARBON-DIOXIDE DETECTOR DURING STABILIZATION AND TRANSPORT OF CRITICALLY ILL CHILDREN

被引:0
作者
BHENDE, MS
THOMPSON, AE
ORR, RA
机构
[1] UNIV PITTSBURGH,SCH MED,DEPT PEDIAT,PITTSBURGH,PA 15261
[2] UNIV PITTSBURGH,SCH MED,DEPT ANESTHESIOL CRIT CARE MED,PITTSBURGH,PA 15261
[3] CHILDRENS HOSP,PEDIAT INTENS CARE UNIT,PITTSBURGH,PA 15213
关键词
END-TIDAL CARBON DIOXIDE DETECTOR; PEDIATRIC TRANSPORT; ENDOTRACHEAL INTUBATION; CRITICALLY ILL CHILDREN; ESOPHAGEAL INTUBATION;
D O I
暂无
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Critically ill children often require endotracheal intubation prior to transport to a medical center. Correct endotracheal tube placement and maintenance during transport are essential. The utility of a portable colorimetric end-tidal CO2 detector during transport of critically ill children was evaluated. Fifty-eight children with spontaneous circulation (aged 1 day to 12 years, weight 0.9 to 26 kg) who underwent 59 intubations during transport by ground (n = 31) or air (n = 27) were studied. Tube position was confirmed by physical examination, arterial blood gas values, or arterial oxygen saturation, and sometimes by chest radiograph. The detector was attached and readings were obtained after intubation; readings were repeated if endotracheal tube position was rechecked during transport. Fifty-seven of 58 tracheal positions and the 1 esophageal tube position were correctly identified. One false-negative result occurred in a severely hypocarbic 900-g premature newborn. On each occasion that the detector was used en route, the endotracheal tube position was correctly identified. It is concluded that the end-tidal CO2 detector is a useful tool for confirming endotracheal tube position during transport of critically ill children weighing more than 2 kg who are not in cardiopulmonary arrest.
引用
收藏
页码:1042 / 1044
页数:3
相关论文
共 19 条
[1]   VALIDITY OF A DISPOSABLE END-TIDAL CARBON-DIOXIDE DETECTOR IN VERIFYING ENDOTRACHEAL-TUBE POSITION IN PIGLETS [J].
BHENDE, MS ;
THOMPSON, AE ;
HOWLAND, DF .
CRITICAL CARE MEDICINE, 1991, 19 (04) :566-568
[2]   VALIDITY OF A DISPOSABLE END-TIDAL CO2 DETECTOR IN VERIFYING ENDOTRACHEAL-TUBE PLACEMENT IN INFANTS AND CHILDREN [J].
BHENDE, MS ;
THOMPSON, AE ;
COOK, DR ;
SAVILLE, AL .
ANNALS OF EMERGENCY MEDICINE, 1992, 21 (02) :142-145
[3]  
BIRMINGHAM PK, 1986, ANESTH ANALG, V65, P886
[4]   STANDARD OF CARE AND ANESTHESIA LIABILITY [J].
CHENEY, FW ;
POSNER, K ;
CAPLAN, RA ;
WARD, RJ .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1989, 261 (11) :1599-1603
[5]  
Gravenstein JHS, 1989, CAPNOGRAPHY CLIN PRA, P43
[6]   END-TIDAL CO2 DETECTION [J].
HELLER, MB ;
YEALY, DM ;
SEABERG, DC ;
MACLEOD, BA ;
GERARD, J .
ANNALS OF EMERGENCY MEDICINE, 1989, 18 (12) :1375-1375
[7]  
JONES B R, 1989, Anesthesiology (Hagerstown), V71, pA359, DOI 10.1097/00000542-198909001-00359
[8]  
KANTER RK, 1989, PEDIATRICS, V84, P43
[9]   CAPNOGRAPHY FOR DETECTION OF ACCIDENTAL ESOPHAGEAL INTUBATION [J].
LINKO, K ;
PALOHEIMO, M ;
TAMMISTO, T .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1983, 27 (03) :199-202
[10]   VERIFICATION OF ENDOTRACHEAL-TUBE PLACEMENT WITH COLORIMETRIC END-TIDAL CO2 DETECTION [J].
MACLEOD, BA ;
HELLER, MB ;
GERARD, J ;
YEALY, DM ;
MENEGAZZI, JJ .
ANNALS OF EMERGENCY MEDICINE, 1991, 20 (03) :267-270