CROSS CALIBRATION OF QDR-2000 AND QDR-1000 DUAL-ENERGY X-RAY DENSITOMETERS FOR BONE-MINERAL AND SOFT-TISSUE MEASUREMENTS

被引:105
作者
ABRAHAMSEN, B
GRAM, J
HANSEN, TB
BECKNIELSEN, H
机构
[1] Department of Medical Endocrinology M, Odense University Hospital, Odense
关键词
BONE; DENSITOMETRY; DUAL-ENERGY X-RAY ABSORPTIOMETRY; BONE MINERAL DENSITY; LONGITUDINAL STUDIES; PRECISION;
D O I
10.1016/8756-3282(94)00054-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Replacement of dual energy X-ray densitometry (DXA) equipment may be necessary in many labs with time, or new equipment may be added. In this context we compared patient scans between a Hologic QDR-1000W and a QDR-2000 (n = 29-43, depending on anatomic region) and between QDR-2000 single beam (SE) and fan beam (FB) modes (n = 40-62) as a quality control measure. A total of 83 subjects (79 females and four males) with a wide range of bone mineral densities (BMD) were studied. There was a linear relationship between results with the QDR-1000W and QDR-2000 in SE mode, and between SE and FB mode on the QDR-2000, but the magnitude of the coefficients and constants differed for the different anatomic regions. In SE mode the QDR-2000 underestimated whole body and forearm BMD by 3% relative to the QDR-1000W, even after cross calibration using a spine phantom. Femoral total BMD was slightly, but not significantly, underestimated. Thus, additional postanalysis correction had to be applied to forearm and whole-body scans. In FB with the QDR-2000, spine and. whole-body BMD was underestimated by 3%, but femur total and neck BMD was overestimated by 2.2 and 2.8%, respectively, compared with SE scans on the same device. Soft-tissue composition with FB (enhanced analysis protocol) on the QDR-2000 differed greatly from that obtained using SE (standard protocol), Lean tissue mass was 4 kg lower and fat mass 4 kg higher in FB mode. Also, total body BMC was 7% lower using FB. Patients should be followed using the same scanning mode and analysis protocol whenever possible. Cross calibration by means of a single BMD anthropomorphic spine phantom is not sufficient to ensure identical results in all anatomic regions when a QDR-2000 replaces a QDR-1000W or is switched to FB mode. Linear regressions in vivo should be determined for each anatomic region.
引用
收藏
页码:385 / 390
页数:6
相关论文
共 10 条
[1]   DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS [J].
BLAKE, GM ;
PARKER, JC ;
BUXTON, FMA ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :902-906
[2]   X-RAY DUAL-PHOTON ABSORPTIOMETRY - A NEW METHOD FOR THE MEASUREMENT OF BONE-DENSITY [J].
CULLUM, ID ;
ELL, PJ ;
RYDER, JP .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (739) :587-592
[3]  
DRINKWATER DT, 1993, J BONE MINER RES, V8, pS352
[4]  
DUNN W L, 1992, Journal of Nuclear Medicine, V33, P1063
[5]  
EIKEN P, 1994, IN PRESS BONE
[6]   CROSS-CALIBRATION OF DXA EQUIPMENT - UPGRADING FROM A HOLOGIC QDR 1000/W TO A QDR-2000 [J].
FAULKNER, KG ;
GLUER, CC ;
ESTILO, M ;
GENANT, HK .
CALCIFIED TISSUE INTERNATIONAL, 1993, 52 (02) :79-84
[7]  
HARPER KD, 1992, J BONE MINER RES, V7, pS19
[8]  
Lilley J, 1991, Osteoporos Int, V1, P141, DOI 10.1007/BF01625443
[9]  
ORWOLL ES, 1991, J BONE MINER RES, V6, P191
[10]  
PRITCHARD JE, 1993, EUR J CLIN NUTR, V47, P216