Accuracy of Preoperative Ultrasonography for Cubital Tunnel Syndrome: A Comparison with Intraoperative Findings

被引:10
作者
Cho, Chul-Hyun [1 ]
Lee, Yong-Ho [1 ]
Song, Kwang-Soon [1 ]
Lee, Kyung-Jae [1 ]
Lee, Si-Wook [1 ]
Lee, Sung-Moon [1 ]
机构
[1] Keimyung Univ, Sch Med, Dongsan Med Ctr, Dept Orthoped Surg, 56 Dalseong Ro, Daegu 41931, South Korea
基金
新加坡国家研究基金会;
关键词
Ulnar nerve; Cubital tunnel syndrome; Ultrasonography;
D O I
10.4055/cios.2018.10.3.352
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The aim of this study was to assess the consistency between preoperative ultrasonographic and intraoperative measurements of the ulnar nerve in patients with cubital tunnel syndrome. Methods: Twenty-six cases who underwent anterior transposition of the ulnar nerve for cubital tunnel syndrome were enrolled prospectively. On preoperative ultrasonography, largest cross-sectional diameters of the ulnar nerve were measured at the level of medial epicondyle (ME) and 3 cm proximal (PME) and distal (DME) to the ME on the transverse scan by a single experienced radiologist. Intraoperative direct measurements of the largest diameter at the same locations were performed by a single surgeon without knowledge of the preoperative values. The consistency between ultrasonographic and intraoperative values including the largest diameter and swelling ratio were assessed. Results: Significant differences between ultrasonographic and intraoperative values of the largest diameter were found at all levels. The mean difference was 1.29 mm for PME, 1.38 mm for ME, and 1.12 mm for DME. The mean ME-PME swelling ratio for ultrasonographic and intraoperative measurements was 1.50 and 1.39, respectively, showing significant difference. The mean ME-DME swelling ratio for ultrasonographic and intraoperative measurements was 1.53 and 1.43, respectively, showing no significant difference. Conclusions: Ultrasonographically measured largest diameters of the ulnar nerve at any levels were smaller than the real values determined intraoperatively. The ME-DME swelling ratio of the ulnar nerve measured by ultrasonography was consistent with the intraoperative measurement.
引用
收藏
页码:352 / 357
页数:6
相关论文
共 17 条
[1]   Ultrasound imaging of the ulnar nerve: Correlation of preoperative and intraoperative dimensions [J].
Bartels, Ronald H. M. A. ;
Meulstee, Jan ;
Verhagen, Wim I. M. ;
Luttikhuis, Thea T. M. Claessen-Oude .
CLINICAL NEUROLOGY AND NEUROSURGERY, 2008, 110 (07) :687-690
[2]   ULTRASONOGRAPHY IN PATIENTS WITH ULNAR NEUROPATHY AT THE ELBOW: COMPARISON OF CROSS-SECTIONAL AREA AND SWELLING RATIO WITH ELECTROPHYSIOLOGICAL SEVERITY [J].
Bayrak, Ayse Oytun ;
Bayrak, Ilkay Koray ;
Turker, Hande ;
Elmali, Muzaffer ;
Nural, Mehmet Selim .
MUSCLE & NERVE, 2010, 41 (05) :661-666
[3]   ULTRASONOGRAPHY IN ULNAR NEUROPATHY AT THE ELBOW: A CRITICAL REVIEW [J].
Beekman, Roy ;
Visser, Leo H. ;
Verhagen, Wim Im .
MUSCLE & NERVE, 2011, 43 (05) :627-635
[4]  
Campbell WW, 1999, MUSCLE NERVE, V22, P408
[5]  
Chiou HE, 1998, J ULTRAS MED, V17, P643
[6]   THE VALIDITY OF ULTRASONOGRAPHIC ASSESSMENT IN CUBITAL TUNNEL SYNDROME: THE VALUE OF A CUBITAL-TO-HUMERAL NERVE AREA RATIO (CHR) COMBINED WITH MORPHOLOGIC FEATURES [J].
Gruber, Hannes ;
Glodny, Bernhard ;
Peer, Siegfried .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2010, 36 (03) :376-382
[7]  
Jeon IH, 2007, J KOREAN SHOULDER EL, V10, P99
[8]   Ultrasonography in ulnar neuropathy at the elbow: Relationships to clinical and electrophysiological findings [J].
Mondelli, M. ;
Filippou, G. ;
Frediani, B. ;
Aretini, A. .
NEUROPHYSIOLOGIE CLINIQUE-CLINICAL NEUROPHYSIOLOGY, 2008, 38 (04) :217-226
[9]   Diagnostic ultrasonography of the ulnar nerve in cubital tunnel syndrome [J].
Okamoto, M ;
Abe, M ;
Shirai, H ;
Ueda, N .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2000, 25B (05) :499-502
[10]   Morphology and dynamics of the ulnar nerve in the cubital tunnel - Observation by ultrasonography [J].
Okamoto, M ;
Abe, M ;
Shirai, H ;
Ueda, N .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2000, 25B (01) :85-89