Reproducibility of Central Corneal Thickness Measurements in Normal Eyes Using the Zeiss Cirrus 5000 HD-OCT and Pentacam HR

被引:13
作者
Baghdasaryan, Elmira [1 ]
Huang, Xiwen [1 ]
Marion, Kenneth M. [1 ]
Tepelus, Tudor C. [1 ]
Bagherinia, Homayoun [3 ]
Sadda, SriniVas R. [1 ,2 ]
Hsu, Hugo Y. [1 ,2 ]
机构
[1] Doheny Eye Inst, 1355 San Pablo St, Los Angeles, CA 90033 USA
[2] Univ Calif Los Angeles, David Geffen Sch Med, Dept Ophthalmol, Los Angeles, CA 90095 USA
[3] Carl Zeiss Meditec Inc, Dublin, CA USA
关键词
Cornea; Central corneal thickness; Imaging; Optical coherency tomography; Pachymetry; Scheimpflug technology;
D O I
10.2174/1874364101812010072
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objectives: To determine the repeatability and reproducibility of Central Corneal Thickness (CCT) measurements using two different anterior segment imaging modalities, including those obtained with the new anterior segment lens attachments for the Cirrus 5000 HD-OCT. Methods: A total of 32 eyes from 16 normal volunteers (8 male, 8 female) were enrolled in this prospective study. CCT was measured by the same examiner using the Cirrus 5000 HD-OCT and Pentacam HR. The results of CCT obtained by each method were averaged and compared using t-test analysis. The agreement between the measurement methods was evaluated. Coefficient of Repeatability (CoR) and Intra-Class Correlation Coefficient (ICC) were computed. Results: The mean measurements taken with the Cirrus OCT anterior chamber lens (CCTAC), HD cornea lens (CCTHDC) and pachymetry scans (CCTPach) were 545.35 +/- 31.02, 537.87 +/- 26.82, and 532.04 +/- 29.82 mu m, respectively. The mean CCT obtained with the Pentacam (CCTPent) was 545.51 +/- 30.71 mu m. CCTPent were significantly higher than CCTHDC and CCTPach (p<0.0001). In contrast, the CCTPent and CCTAC were similar (p=0.87). CCT, as evaluated by the two different instruments, showed excellent correlation (r > 0.98, p<0.0001) with an ICC > 0.99 (95% CI, 0.97 - 0.99). CoR was the highest for CCTPach (3.7 +/- 1.4, 95% CI (3.0-4.6)). Conclusion: CCT measurements from the Cirrus OCT using the new anterior segment lens attachments and the Pentacam HR are highly correlated. This should allow the use of a standardized correction factor if necessary to inter-relate the measurements between the two devices.
引用
收藏
页码:72 / 83
页数:12
相关论文
共 36 条
[1]   Corneal-thickness spatial profile and corneal-volume distribution: Tomographic indices to detect keratoconus [J].
Ambrosio, Renato, Jr. ;
Alonso, Ruiz Simonato ;
Luz, Allan ;
Coca Velarde, Luis Guillermo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (11) :1851-1859
[2]   Corneal ectasia after laser in situ keratomileusis [J].
Argento, C ;
Cosentino, MJ ;
Tytiun, A ;
Rapetti, G ;
Zarate, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (09) :1440-1448
[3]   Collagen cross-linkage: a comprehensive review and directions for future research [J].
Ashwin, P. T. ;
McDonnell, P. J. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (08) :965-970
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Central corneal thickness and measured IOP response to topical ocular hypotensive medication in the ocular hypertension treatment study [J].
Brandt, JD ;
Beiser, JA ;
Gordon, MO ;
Kass, MA .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 138 (05) :717-722
[6]   Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes [J].
Chakrabarti, HS ;
Craig, JP ;
Brahma, A ;
Malik, TY ;
McGhee, CNJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) :1823-1828
[7]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[8]   Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation [J].
Doors, Muriel ;
Cruysberg, Lars P. J. ;
Berendschot, Tos T. J. M. ;
de Brabander, John ;
Verbakel, Frenne ;
Webers, Carroll A. B. ;
Nuijts, Rudy M. M. A. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2009, 247 (08) :1139-1146
[9]   Diurnal variation of corneal sensitivity and thickness [J].
du Toit, R ;
Vega, JA ;
Fonn, D ;
Simpson, T .
CORNEA, 2003, 22 (03) :205-209
[10]   Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound [J].
Gherghel, D ;
Hosking, SL ;
Mantry, S ;
Banerjee, S ;
Naroo, SA ;
Shah, S .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (06) :1272-1277