HIGH-SPEED PHOTOGRAPHIC EVALUATION OF ENDOSCOPIC LITHOTRIPSY DEVICES

被引:9
作者
STERENBORG, HJCM
DEREIJKE, TM
WIERSMA, J
ERCKENS, RC
JOGSMA, FHM
机构
[1] UNIV AMSTERDAM, ACAD HOSP, DEPT UROL, 1105 AZ AMSTERDAM, NETHERLANDS
[2] UNIV LIMBURG, DEPT OPHTHALMOL, 6200 MD MAASTRICHT, NETHERLANDS
来源
UROLOGICAL RESEARCH | 1991年 / 19卷 / 06期
关键词
HIGH-SPEED PHOTOGRAPHY; SHOCK WAVE; ELECTROHYDRAULIC LITHOTRIPSY; LASER LITHOTRIPSY; STONE FRAGMENTATION;
D O I
10.1007/BF00310154
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Shock-wave generation and bubble formation occurring during endoscopic lithotripsy were studied using high-speed photography for various devices: a pulsed-dye laser, a Q-switched Nd:YAG laser and an electrohydraulic (EHL) apparatus. The three devices investigated generated gas bubbles that rapidly expanded and decayed. The maximal size of these bubbles was 5-8 mm for the pulsed-dye laser, 6 mm for the Q-switched Nd: YAG laser and up to 14 mm for the EHL device. The bubble size appeared to be governed mainly by the energy per pulse delivered by the lithotripsy device. The shock-wave pattern depended strongly on the type of device used; the 25-ns Q-switched Nd:YAG laser pulse generated a single pressure step, whereas the 1.5-mu-s pulsed-dye laser produced a train of shock fronts.