EVALUATION OF KEELER PULSAIR NONCONTACT TONOMETRY - RELIABILITY AND REPRODUCIBILITY

被引:15
作者
BONOMI, L [1 ]
BARAVELLI, S [1 ]
COBBE, C [1 ]
TOMAZZOLI, L [1 ]
机构
[1] UNIV VERONA,INST OPHTHALMOL,CTR OSPED BG TRENTO,I-37126 VERONA,ITALY
关键词
D O I
10.1007/BF00167869
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Intraocular pressure was evaluated in 414 eyes using Keeler Pulsair tonometry in comparison with values obtained by Goldmann tonometry. No statistically significant differences were found between the two techniques. Landis K statistical analysis was used to evaluate the measurements obtained by two operators in 125 eyes, with the results indicating good reproductibility. In 20 eyes, a statistically significant difference was found between values obtained with patients in the seated vs supine positions, whereas measurements made with subjects' heads in a supported or unsupported position yielded no statistically different values.
引用
收藏
页码:210 / 212
页数:3
相关论文
共 11 条
  • [1] BUSCEMI M, 1989, Annali di Ottalmologia e Clinica Oculistica, V115, P309
  • [2] COBBE C, 1990, IN PRESS B OCUL, V69
  • [3] DIBARI A, 1988, B OCUL, V67, P157
  • [4] DUCREY N, 1975, OPHTHALMOLOGICA, V173, P446
  • [5] A NEW HAND-HELD AIR IMPULSE TONOMETER
    FISHER, JH
    WATSON, PG
    SPAETH, G
    [J]. EYE, 1988, 2 : 238 - 242
  • [6] NONCONTACT APPLANATION TONOMETER - DESCRIPTION AND CLINICAL EVALUATION
    FORBES, M
    PICO, G
    GROLMAN, B
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1974, 91 (02) : 134 - 140
  • [7] INSLER MS, 1987, ARCH OPHTHALMOL-CHIC, V105, P1358
  • [8] SORENSEN PN, 1975, ACTA OPHTHALMOL, V53, P513
  • [9] SPONSEL WE, 1989, ACTA OPHTHALMOL, V67, P567
  • [10] NON-CONTACT TONOMETRY IN THE POSTOPERATIVE EYE
    VERNON, SA
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 1989, 73 (04) : 247 - 249