CONSISTENT SUBCUTANEOUS PREPECTORAL IMPLANTATION OF A NEW IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR

被引:41
作者
STANTON, MS
HAYES, DL
MUNGER, TM
TRUSTY, JM
ESPINOSA, RE
SHEN, WK
OSBORN, MJ
PACKER, DL
HAMMILL, SC
机构
[1] Division of Cardiovascular Diseases and Internal Medicine, Mayo Clinic Rochester, Rochester, Minnesota
关键词
defibrillation threshold; DFT; ICD; implantable cardioverter defibrillator; superior vena cava; SVC;
D O I
10.1016/S0025-6196(12)62213-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To describe the use of a new implantable cardioverter defibrillator (ICD) that can be placed in the prepectoral region rather than implanted in the abdominal wall. Design: We report the experience of placement of this new ICD in the prepectoral region in 13 patients from Sept. 28, 1993, through Jan. 10, 1994, at the Mayo Clinic. Material and Methods: Thirteen consecutive patients offered this new ICD underwent placement of transvenous defibrillation leads, and the pulse generator was placed in a pocket formed in the subcutaneous, prepectoral space. Testing ensured a defibrillation threshold of 24 J or less. Results: In all 13 patients, the pulse generator could be placed in the subcutaneous, prepectoral space. In all except one patient, acceptable defibrillation thresholds were achieved by using lead systems placed totally transvenously. Only one patient required placement of a subcutaneous patch. All but two patients were dismissed from the hospital within 3 days after the ICD implantation. Conclusion: Consistent subcutaneous, prepectoral placement of this new ICD pulse generator is possible. Because the entire procedure can be performed in the pacemaker implantation room, the potential exists for decreasing the duration of the hospitalization and associated costs.
引用
收藏
页码:309 / 314
页数:6
相关论文
共 9 条
  • [1] IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS
    BARDY, GH
    HOFER, B
    JOHNSON, G
    KUDENCHUK, PJ
    POOLE, JE
    DOLACK, GL
    GLEVA, M
    MITCHELL, R
    KELSO, D
    [J]. CIRCULATION, 1993, 87 (04) : 1152 - 1168
  • [2] COMPARATIVE EFFICACY OF MONOPHASIC AND BIPHASIC TRUNCATED EXPONENTIAL SHOCKS FOR NONTHORACOTOMY INTERNAL DEFIBRILLATION IN DOGS
    CHAPMAN, PD
    VETTER, JW
    SOUZA, JJ
    TROUP, PJ
    WETHERBEE, JN
    HOFFMANN, RG
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (03) : 739 - 745
  • [3] GUIDELINES FOR IMPLANTATION OF CARDIAC-PACEMAKERS AND ANTIARRHYTHMIA DEVICES - A REPORT OF THE AMERICAN-COLLEGE-OF-CARDIOLOGY AMERICAN-HEART-ASSOCIATION TASK-FORCE ON ASSESSMENT OF DIAGNOSTIC AND THERAPEUTIC CARDIOVASCULAR PROCEDURES (COMMITTEE ON PACEMAKER IMPLANTATION)
    DREIFUS, LS
    FISCH, C
    GRIFFIN, JC
    GILLETTE, PC
    MASON, JW
    PARSONNET, V
    [J]. CIRCULATION, 1991, 84 (01) : 455 - 467
  • [4] SUBPECTORAL IMPLANTATION OF CARDIOVERTER-DEFIBRILLATOR COMBINED WITH A NONEPICARDIAL LEAD SYSTEM - PRELIMINARY EXPERIENCE WITH A NOVEL-APPROACH
    IP, JH
    MEHTA, D
    PE, E
    CAMUNAS, JL
    GOMES, JA
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (11) : 857 - 860
  • [5] COMPARISON OF THE INTERNAL DEFIBRILLATION THRESHOLDS FOR MONOPHASIC AND DOUBLE AND SINGLE CAPACITOR BIPHASIC WAVEFORMS
    KAVANAGH, KM
    TANG, ASL
    ROLLINS, DL
    SMITH, WM
    IDEKER, RE
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (05) : 1343 - 1349
  • [6] A POPULATION-BASED ESTIMATE OF CANDIDACY RATES FOR THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    KOTTKE, TE
    STANTON, MS
    BAILEY, KR
    DECKER, WW
    HAMMILL, SC
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (01) : 77 - 81
  • [7] AN ANALYSIS OF THE COST-EFFECTIVENESS OF THE IMPLANTABLE DEFIBRILLATOR
    KUPPERMANN, M
    LUCE, BR
    MCGOVERN, B
    PODRID, PJ
    BIGGER, JT
    RUSKIN, JN
    [J]. CIRCULATION, 1990, 81 (01) : 91 - 100
  • [8] AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - IS EARLY IMPLANTATION COST-EFFECTIVE
    ODONOGHUE, S
    PLATIA, EV
    BROOKSROBINSON, S
    MISPIRETA, L
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (05) : 1258 - 1263
  • [9] DEFIBRILLATION ENERGY-REQUIREMENTS WITH SINGLE ENDOCARDIAL (ENDOTAK(TM)) LEAD
    WINTER, J
    VESTER, EG
    KUHLS, S
    KANTARTZIS, M
    PERINGS, C
    PAUSCHINGER, M
    STRAUER, BE
    BIRCKS, W
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 540 - 546